COMPARISON OF AUTOMATIC STARRSED RS ERYTHROCYTE SEDIMENTATION RATE WITH MANUAL WESTERGREN METHOD Ninik Sukartini; Siskawati Suparmin
INDONESIAN JOURNAL OF CLINICAL PATHOLOGY AND MEDICAL LABORATORY Vol 25, No 3 (2019)
Publisher : Indonesian Association of Clinical Pathologist and Medical laboratory

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.24293/ijcpml.v25i3.1302

Abstract

This study aimed to compare the results of ESR measurements using Starrsed RS (30- and 60-minute mode) and manual Westergren method. We also investigate the effect of high leukocytes and haematocrit (Ht) on Starrsed RS ESR measurements. A cross-sectional study using K3EDTA anticoagulated blood samples from 140 subjects were conducted. A total of 21 samples with leukocytes count >50 000/µl were examined. The effect of Ht was studied using data from comparative study subjects.  The 60-minute mode Starrsed RS correlated very strongly [r = 0.98 (95%CI 0.97 to 0.98), p <0.0001], had good agreement [mean bias -0.4 mm (95%CI -1.46 to 0.74)], and gave no significant different results with manual Westergren method. The 30-minute mode Starrsed RS also has a very strong correlation [r = 0.97 (95%CI 0.96 to 0.98), p<0.0001] and good agreement [mean bias-2.9 mm (95%CI -4.23 to -1.58)], but has systematic difference with manual Westergren method. There was significant difference between manual method and Starrsed RS in samples with high leukocytes (p = 0.0208). There were no significant differences of ESR results measured by Starrsed RS dan manual Westergren method in the group of Ht≤35% and >35%. The ESR on the Starrsed RS should still be read at 60 minute. Leukocytes>50 000/µl may affect ESR measurements on Starrsed but no effect of haematocrit was observed.
Evaluasi Pemeriksaan Imunoglobulin E Spesifik Menggunakan Immunoblot Assay dengan Baku Emas Skin Prick Test Yudhistira -; Ninik Sukartini; Suzanna Immanuel; Iris Rengganis
Cermin Dunia Kedokteran Vol 46, No 2 (2019): Penyakit Dalam
Publisher : PT. Kalbe Farma Tbk.

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.55175/cdk.v46i2.517

Abstract

Pendahuluan. Skin Prick Test (SPT) merupakan baku emas diagnosis alergi, tetapi tidak dapat dilakukan pada kondisi tertentu seperti dermatografisme, hamil, tidak dapat lepas obat antialergi, sehingga pemeriksaan IgE spesifik menjadi pilihan. Di Indonesia belum ada data perbandingan pemeriksaan IgE spesifik dengan SPT. Penelitian ini dilakukan untuk menentukan sensitivitas, spesifisitas, PPV, NPV, LR+ dan LR- pemeriksaan IgE spesifik menggunakan analisis immunoblot. Metode. Penelitian menggunakan desain potong lintang. Subjek penelitian adalah pasien poliklinik alergi imunologi. Pengambilan sampel dengan metode non-probability sampling dengan teknik consecutive sampling. Analisis dilakukan terhadap alergen tungau debu rumah (Dermatophagoides pteronyssinus, Dermatophagoides farinae, Blomia tropicalis) dan kecoa. Hasil. Seratus subjek yang memenuhi kiteria masukan dan tolakan (76% perempuan, rerata usia 38.9 tahun) ikut serta dalam penelitian ini. Sensitivitas empat alergen bervariasi dengan rentang 32.4%-76.8%, spesifisitas 68.0%-85.7%, PPV 54.5%-94.5%, NPV 46.2%-65.3%, LR+ 1.8-5.0, dan LR- 0.3-0.8.Sensitivitas pemeriksaan IgE spesifik cukup baik pada tiga tungau debu rumah tetapi rendah pada kecoa; spesifisitas dan PPV bervariasi cukup sampai baik; NPV cukup baik. Simpulan dan Saran. Uji diagnostik IgE spesifik tungau debu rumah menunjukkan hasil cukup baik. Pemeriksaan IgE spesifik tidak dapat digunakan untuk skrining alergi kecoa.Background. Skin Prick Test (SPT) is considered as gold standard for diagnosis of allergy, but cannot be performed in some conditions such as dermatographism, pregnancy, and unable to abstain from antiallergy drugs. In these conditions, specific-IgE test is a test of choice. But there is no comparison data on specific IgE test with SPT in Indonesia. This study was to investigate sensitivity, specificity, PPV, NPV, LR+, and LR- of specific-IgE test of immunoblot assay kit. Method. The study was cross-sectional. Subjects were patients from allergy-immunology clinic who fulfilled inclusion and exclusion criteria. Sampling was performed using non-probability sampling method with consecutive technique. Analysis was performed for house dust mite allergen (Dermatophagoides pteronyssinus, Dermatophagoides farinae, Blomia tropicalis) and cockroach. Result. Hundred subjects (76% was female, average age of 38.9 years) were involved. Sensitivity of four allergens were varied with range 32.4%-76.8%, specificity 68.0%-85.7%, PPV 54.5%-94.5%, NPV 46.2%-65.3%, LR+ 1.8-5.0, and LR- 0.3-0.8. Sensitivities of specific-IgE for three species of house dust mite were moderate, but low for cockroach; specificities and PPVs were moderate to high; NPVs were moderate. Conclusion. Diagnostic test of specific-IgE with house dust mite allergen shows adequate result. Specific IgE test cannot be utilized for screening of cockroach allergy. 
Comparison of the Profile and TSH Levels from Several Types of Blood Collection Tubes Gunawan Eka Putra; Ninik Sukartini; Suzanna Immanuel; Fify Henrika; Nuri Dyah Indrasari
INDONESIAN JOURNAL OF CLINICAL PATHOLOGY AND MEDICAL LABORATORY Vol. 26 No. 2 (2020)
Publisher : Indonesian Association of Clinical Pathologist and Medical laboratory

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.24293/ijcpml.v26i2.1475

Abstract

Thyroid-Stimulating Hormone (TSH) is an important parameter in diagnosing thyroid disease which uses serum according to the World Health Organization's (WHO) recommendations. The use of plasma can help improve the Turn Around Time (TAT); however, the discrepancy with serum is unknown. A cross-sectional study using 89 blood samples was performed to compare TSH levels using serum tubes with clot activator (Tube I), plasma tubes with heparin (Tube II), and plasma tubes with heparin-gel separator (Tube III); and to overview of TSH levels according to gender and age. The median of TSH levels in Tubes I, II, and III were 1.380 (0.032-7.420) μIU/mL, 1.380 (0.030-7.480) μIU/mL, and 1.360 (0.030-7.460) μIU/mL, respectively. There were no statistically significant differences in TSH levels of the three tubes. The median TSH levels differences of Tubes II and III compared to the tube I were -0.9% (-7.2-2.2) and -1.7% (-8.0-1.6), respectively. Measurement bias observed in this study was following the specified desirable bias according to Ricos. The median TSH levels of the male and female groups were 1.500 (0.032-4.250) μIU/mL and 1.345 (0.058-7.420) μIU/mL, respectively. Median TSH levels of 31-40 years old age group and >61 years old age group were 1.190 (0.609-3.240) μIU/mL and 1.730 (0.088-5.760) μIU/mL, respectively. Specimens from three tubes could be used to examine TSH levels. Measurement of TSH levels showed a higher median in the male and older group.
Amino Acid Profile of Luminal A and B Subtypes Breast Cancer Sonar Soni Panigoro; Arif Kurniawan; Ramadhan Ramadhan; Ninik Sukartini; Herqutanto Herqutanto; Rafika Indah Paramita; Ferry Sandra
The Indonesian Biomedical Journal Vol 15, No 3 (2023)
Publisher : The Prodia Education and Research Institute (PERI)

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.18585/inabj.v15i3.2109

Abstract

BACKGROUND: Amino acids are important for proliferation and maintenance of tumor cells. Breast cancer patients were found to have significant changes in the number of amino acids, which are assumed to be correlated with the molecular subtypes of breast cancer. Therefore, current study was conducted to analyze plasma amino acids in breast cancer patients with luminal A and B subtypes.METHODS: Breast cancer and control subjects were recruited, and venous blood was collected for the measurement of plasma amino acids. Total 19 plasma amino acids were measured using reverse-phase high-performance liquid chromatography with C18 column. Mean comparison for normally distributed and homogeneous data was further analzyed using independent sample T-test, with p<0.05 was considered as significant.RESULTS: From total 19 amino acids, only 7 amino acids; cysteine, glutamic acid, histidine, ornithine, threonine, tyrosine, valine, were statistically different between the healthy control and breast cancer subjects. Eventhough no amino acids was found to be statistically different between breast cancer subjects with luminal A and B subtypes, but some amino acids were found to be significantly different when correlated to various breast cancer risk factors.CONCLUSION: Amino acid profile of patients with Luminal A and B subtypes of breast cancer differs compared to healthy controls and is also correlated with breast cancer risk factors. Increase in cysteine level in Luminal A subtype patients and decrease of alanine and leucine in Luminal B subtype patients can be used as a biomarker.KEYWORDS: amino acid, plasma, breast cancer, risk factor, biomarker
The Role of Liquid Biopsy for Detecting Epidermal Growth Factor Receptor (EGFR) Mutations in Patients with Advanced Non-Small Cell Lung Cancer (NSCLC) – A Preliminary Study in Indonesian Population Agus Susanto Kosasih; Arif Riswahyudi Hanafi; Ninik Sukartini; Mariska Pangaribuan; Muhammad Alfin Hanif; Willy Pandu Ariawan; Sri Agustini Kurniawati; Dian Cahyanti; Kartika Anastasia Kosasih; Alyssa Diandra; Markus Yovian; Lyana Setiawan
Indonesian Journal of Cancer Vol 17, No 3 (2023): September
Publisher : http://dharmais.co.id/

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.33371/ijoc.v17i3.1061

Abstract

Background: The development of new biomarker-targeted therapies is associated with the improvements in overall survival of advanced non-small cell lung cancer (NSCLC) sufferers. However, tissue biopsy as the gold standard to determine tumor molecular status is not always feasible in advanced diseases. Liquid biopsy, a minimally invasive technique to obtain cytological and molecular assessment from patients allows real-time monitoring of tumors and identification of resistant mechanisms. Therefore, this preliminary study aimed to compare the molecular profile of advanced NSCLC from liquid and tissue biopsy.Methods:  This cross-sectional study was conducted on patients with NSCLC undergoing diagnostic procedures at Dharmais National Cancer Center Hospital from January 2018 to December 2021. Tissue biopsy to check epidermal growth factor receptor (EGFR) mutations was performed on formalin-fixed paraffin-embedded (FFPE) tissue blocks. Analysis of mutations involved using a home-brew polymerase chain reaction (PCR) based on a high-resolution melting technique. Meanwhile, the liquid biopsy was performed using blood samples, and sequencing for circulating tumor DNA (ctDNA) was carried out with NGS.Results: : The results showed that among the 22 subjects enrolled in the study, tissue biopsy identified wild-type EGFR in 18 (81.8%), exon 19 mutations in 3, and exon 21 mutations (L861Q and L858R) in 3. Liquid biopsy found EGFR exon 19 deletions in 11 samples (50%), exon 21 mutations (L861Q and L858R) in 10, L861Q in 7, L8585 in 6, and wild type in 1 sample. Additionally, acquired EGFR mutations in exon 20 T790M were found in 5 samples.Conclusions: Liquid biopsy may be beneficial in patients for whom tissue biopsy cannot be performed or where tissue for molecular profiling is inadequate. This modality can also be used to detect molecular resistance in patients with advanced NSCLC.
A Rare Case: HaemoglobinS-Thalassemia in Adult Linny Luciana; Ninik Sukartini
Jurnal Kedokteran Meditek Vol 29 No 2 (2023): MEI
Publisher : Fakultas Kedokteran Universitas Kristen Krida Wacana

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.36452/jkdoktmeditek.v29i2.2540

Abstract

Hemoglobinopathy refers to a disease involving a qualitative or quantitative defect of the structure or synthesis of haemoglobin molecules. The HaemoglobinS- beta thalassemia occurs in a heterozygotes individual with beta-thalassemia and HaemoglobinS gene.  A 29-year-old man came with severe anemia, thrombocytopenia, and history of repeated blood transfusions. Physical examination showed pale conjunctiva, pansystolic murmurs, and hepatosplenomegaly. The HaemoglobinS fraction was found in haemoglobin electrophoresis with increased HaemoglobinF and decreased HaemoglobinA2 fraction. The peripheral blood smear shows abnormal erythrocytes morphologies such as pencil shapes, fragmentocytes, target cells, and sickle shapes. The patient was diagnosed with chronic anaemia caused by HaemoglobinS-beta thalassemia. It makes ineffective erythropoiesis, intravascular, and extravascular hemolysis. This haemoglobinopathy caused increased ferritin and transferrin saturation. The presence of renal failure indicate there is a complicated condition like microvascular obstruction of renal. In this case, there is a reduction of HaemoglobinA2 fraction that is not common in HaemoglobinS-beta thalassemia. The patient with Haemoglobin S / beta+ thalassemia shows intravascular hemolysis, ineffective hematopoiesis, and vaso-occlusive signs. Deoxyribo Nucleic Acid analysis is further needed to confirm the combination defect of haemoglobin synthesis disorders in conjunction with alpha thalassemia or Hereditary persistence of fetal haemoglobin.
Detecting Iron Deficiency Anemia in Type C Hospital: Role of RDW and MCV Parameters Marina Ludong; Ninik Sukartini; Nuri Dyah Indrasari; Dewi Wulandari
INDONESIAN JOURNAL OF CLINICAL PATHOLOGY AND MEDICAL LABORATORY Vol. 30 No. 2 (2024)
Publisher : Indonesian Association of Clinical Pathologist and Medical laboratory

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.24293/ijcpml.v30i2.2100

Abstract

Iron deficiency anemia remains a global health problem, which is also a prominent cause of morbidity and mortality of all range of ages. There are three stages of anemia development, and there are some parameters to detect bodily iron status. Transferrin saturation is one of the reliable parameters. Among all hematology parameters, Red Cell Distribution Width (RDW) and Mean Corpuscular Volume (MCV) are two of the most often studied. MCV and RDW are relatively affordable and accessible, most importantly for rural areas with lower socioeconomic status. This was an analytical observational study with a cross-sectional design aimed to determine the correlation between RDW and MCV values with iron deficiency anemia, which was measured by transferrin saturation. A significant correlation was found between RDW, MCV values, and iron deficiency anemia in patients of Mitra Keluarga Cikarang Hospital and Permata Keluarga Hospital, Jakarta with a p-value of <0.05. Sensitivity and specificity for MCV were 75% and 100%, for RDW were 55.45% and 80%, respectively. In conclusion, RDW and MCV parameters can be used as screening instruments for iron deficiency anemia.
Evaluasi Pemeriksaan Imunoglobulin E Spesifik Menggunakan Immunoblot Assay dengan Baku Emas Skin Prick Test Yudhistira; Ninik Sukartini; Suzanna Immanuel; Iris Rengganis
Cermin Dunia Kedokteran Vol 46 No 2 (2019): Interna
Publisher : PT Kalbe Farma Tbk.

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.55175/cdk.v46i2.505

Abstract

Pendahuluan. Skin Prick Test (SPT) merupakan baku emas diagnosis alergi, tetapi tidak dapat dilakukan pada kondisi tertentu seperti dermatografisme, hamil, tidak dapat lepas obat antialergi, sehingga pemeriksaan IgE spesifik menjadi pilihan. Di Indonesia belum ada data perbandingan pemeriksaan IgE spesifik dengan SPT. Penelitian ini dilakukan untuk menentukan sensitivitas, spesifisitas, PPV, NPV, LR+ dan LR- pemeriksaan IgE spesifik menggunakan analisis immunoblot. Metode. Penelitian menggunakan desain potong lintang. Subjek penelitian adalah pasien poliklinik alergi imunologi. Pengambilan sampel dengan metode non-probability sampling dengan teknik consecutive sampling. Analisis dilakukan terhadap alergen tungau debu rumah (Dermatophagoides pteronyssinus, Dermatophagoides farinae, Blomia tropicalis) dan kecoa. Hasil. Seratus subjek yang memenuhi kiteria masukan dan tolakan (76% perempuan, rerata usia 38.9 tahun) ikut serta dalam penelitian ini. Sensitivitas empat alergen bervariasi dengan rentang 32.4%-76.8%, spesifisitas 68.0%-85.7%, PPV 54.5%-94.5%, NPV 46.2%-65.3%, LR+ 1.8-5.0, dan LR- 0.3-0.8.Sensitivitas pemeriksaan IgE spesifik cukup baik pada tiga tungau debu rumah tetapi rendah pada kecoa; spesifisitas dan PPV bervariasi cukup sampai baik; NPV cukup baik. Simpulan dan Saran. Uji diagnostik IgE spesifik tungau debu rumah menunjukkan hasil cukup baik. Pemeriksaan IgE spesifik tidak dapat digunakan untuk skrining alergi kecoa. Background. Skin Prick Test (SPT) is considered as gold standard for diagnosis of allergy, but cannot be performed in some conditions such as dermatographism, pregnancy, and unable to abstain from antiallergy drugs. In these conditions, specific-IgE test is a test of choice. But there is no comparison data on specific IgE test with SPT in Indonesia. This study was to investigate sensitivity, specificity, PPV, NPV, LR+, and LR- of specific-IgE test of immunoblot assay kit. Method. The study was cross-sectional. Subjects were patients from allergy-immunology clinic who fulfilled inclusion and exclusion criteria. Sampling was performed using non-probability sampling method with consecutive technique. Analysis was performed for house dust mite allergen (Dermatophagoides pteronyssinus, Dermatophagoides farinae, Blomia tropicalis) and cockroach. Result. Hundred subjects (76% was female, average age of 38.9 years) were involved. Sensitivity of four allergens were varied with range 32.4%-76.8%, specificity 68.0%-85.7%, PPV 54.5%-94.5%, NPV 46.2%-65.3%, LR+ 1.8-5.0, and LR- 0.3-0.8. Sensitivities of specific-IgE for three species of house dust mite were moderate, but low for cockroach; specificities and PPVs were moderate to high; NPVs were moderate. Conclusion. Diagnostic test of specific-IgE with house dust mite allergen shows adequate result. Specific IgE test cannot be utilized for screening of cockroach allergy.